MARK-1002
Sodium analyzer Is a dual-channel analyzer for continuous measuring of sodium concentration
as CNa (or PNa) and temperature for high purity water environments.
Warranty 12 months
![]() ![]() ![]() ![]() |
Basic kit:
- Converting unit
- Hydraulic panel HP-1002 or HP-1002T
- Power supply unit
- 5 meter connecting cable C1002.5
- Operation manual
Optionally:
- Hydraulic panel and power supply unit for the second channel
- Connecting cable C1002.L up to 100 meters
- Sample collecting kit
- OPC-server
2 channels
Programmable ranges of measurements for each channel.
Independent measurements in two points.
Convenience and accuracy of measurement, minimum maintenance
Measuring range from 0,01 CNa. Measurement accuracy 6%.
Long inter-calibration period (up to 6 months).
Dual automatic temperature compensation.
Automatic batching device of the alkalizing reagent.
Absence of KCL flask.
1 solution calibration
If necessary, 3-point-calibration is possible.
«Active» sensor unit
Sensor cable length up to 100 m.
Communication with external devices
2 galvanic isolated current outputs 0–5/4–20/0–20mA.
RS 485 galvanic isolated port.
Programmable setpoints for each channel.
|
Measuring range |
Resolution |
Accuracy |
|
---|---|---|---|---|
CNa, ppb |
МАRК-1002
МАRК-1002Т
|
0,7-1000
1000-30001
0,01-1000
|
0,1
0,01
|
±(0,5+0,12*CNa)
±0.3*CNa
±(0,03+0,12*CNa)
|
pNa |
МАRК-1002
МАRК-1002Т
|
4,36-7,52 4,06-7,521 |
0,01 | |
Температура, °C |
0-50 |
0,1 |
±0,3 |
|
Converting unit |
Hydraulic panel |
|
---|---|---|---|
Mounting |
Wall |
Panel |
|
Dimensions, mm |
266*170*95 |
252*146*100 |
300*650*200 |
Weight, kg |
2,6 |
4,0 |
|
Power supply |
220V, 50Hz /10V·А |
24 В |
Environment requirements
Temperature, °С |
10-40 |
Sample flow rate, dm³/min |
0,05-3 |
Temperature of ambient air, °С |
5-50 |
Manual МАRК-1002