MARK-602T
Conductivity meter Is a dual-channel meter for continuous measurement of specific electrical conductivity and resistance (absolute and adjusted to 20° C or 25° C), salinity and temperature of water and aqueous solutions, including deionized and high purity water.
Warranty 12 months
![]() ![]() ![]() ![]() |
Basic kit:
- Converting unit
- Amplifier unit
- Conductivity sensor CS-003T or CS-3T
- 5 meter connecting cable C602T.5
- Operation manual
Optionally:
- Conductivity sensor CS-003T or CS-3T for the second channel
- Flow sensor
- Amplifier unit for the second sensor
- Connecting cable C602T.L up to 1000 m
- Hydraulic panel HP-602T
2 channels
Programmable ranges of measurements for each channel.
Independent measurements in two points.
Convenience and accuracy of measurement, minimum maintenance
Dual automatic temperature compensation, selectable linear coefficient of compensation.
Possibility of placing the converting unit on the remote distance from the point of control
Sensor cable length up to 1000 m.
Communication with external devices
2 galvanic isolated current outputs 0–5/4–20/0–20mA.
RS 485 galvanic isolated port.
Programmable setpoints for each channel.
Durable aluminum case IP65
Instrument is protected from dust and moisture.
Graphic LCD display with backlight
Easy input of all parameters by keypad.
|
Sensor |
Measuring range |
Accuracy |
---|---|---|---|
Conductivity, µS/cm |
ДП-3Т ДП-003Т |
0-20 000 |
±(0,03+ 2% of measured value)
±(0,001+ 2% of measured value)
|
Salinity, mg/L |
ДП-3Т |
0-10 000 |
±(0,03+ 2,5% of measured value)
±(0,001+2,5% of measured value)
|
ρ, кΩ/сm |
ДП-3Т |
0,05-1 000 |
±(0,0005+2% of measured value)
±(0,05+2% of measured value)
|
Temperature, °С |
0-130* |
±0,3 |
|
* temperature compensation range 0-100 °C |
|||
Mounting |
Wall |
Panel |
|
Dimensions, mm |
266*170*95 |
252*146*100 |
|
Weight, kg |
2,60 |
2,60 |
|
|
|||
Power supply |
220V or 36V, 50Hz/15V·А |
Environment requirements
Temperature, °С |
0-130 |
Ambient overpressure, bar, max. |
16 |
DataSheet МАRК-602T