MARK-2010
Portable gas analyzer Occasional monitoring of oxygen share by volume in different gases (hydrogen, natural gas, boiler exhaust gases, nitrogen, argon, helium and others).
Control of hydrogen purity in a generator stator coolant circuit; oxygen monitoring in boiler gas lines during steam raising, at electrolysis units, at working areas, etc.
Warranty 12 months
![]() ![]() |
Basic kit:
- Converting unit
- Oxygen sensor
- Spare parts kit for the sensor
- A power supply unit with a charger
Advantages
Technical specifications
Документация
Stable operation of the oxygen sensor in hydrogen ambient /
Ability to work at small gas flow rates (20cm 3/min)
Measuring in dry and wet media
Automatic temperature and barometric pressure compensation
Unique patented sensor design /
High reaction rate.
Increased mechanical resistance of the construction.
Sealed ultra-strong water-repellent membrane.
Routine maintenance 1 per year /
Calibration against atmospheric oxygen max. 1 per month.
Easy and quick membrane replacement.
Organic high-contrast LED indicator
Durable aluminium case IP65/
Instrument is protected against dust and moisture.
Built-in durable LiFePO4 accumulator /
At least 100 cycles of charge / discharge.
|
Measuring range |
Resolution |
Accuracy |
Oxygen share by volume, % |
0 -30 |
0.001 |
±(0.01+0.04*A) |
A-measured value
|
|||
|
Converting unit |
||
Dimensions, mm |
160*147*172 |
||
Weight, kg |
1 |
||
Environment requirements |
|||
Temperature, C |
1-50 |
||
Pressure, kPa, max. |
7 |
||
Sample flow rate, dm3/min, min. |
0.15-0,4 |